Let us know go into the details of the results which contribute
to 5-th entry of Tab. 1, reported
in Fig. 12 [8]
and in the entries 5-8 of Tab. 2.
Table:
Details of the individual result used in the overall analysis.
|
Figure:
Standard and sceptical combination of the four data points
of Ref. [8]. The dashed line is the simple weighted average;
the gray line having the same center and larger width is the
same with the standard deviation scaled by
as done in Ref. [8]. The thick, asymmetric curve represent
the result of the sceptical combination.
 |
There is one high precision value
favoring a small mass value (
MeV),
and three values of minor precision preferring higher
mass values.
The simple weighted average
of
MeV is then practically equal
to the highest precision value. But then a
scaling is applied by the authors.
The combined uncertainty grows up,
which is something desirable, but
it does it symmetrically around the mean,
not taking into account the fact that the other
results would pull the mass value up.
It is then interesting to make a sceptical combination of these
four points. The result is shown if Fig. 14.
The sceptical analysis takes into account also the results favoring
higher mass values, although the peak of the distribution (the `mode')
remains very close to the most precise result,
and there is a substantial overlap with it.
The distribution is now skewed on the right side, assigning
higher probability that the mass value is, for example, above
MeV
with respect to what we could think judging from
mean and standard deviation alone. The resulting mass is
MeV shifted up by about
keV, with a standard
uncertainty about 50% larger than that provided
by the
scaling prescription.
But what is more interesting is that the latter
(gray line in the figure,
just below the red dashed one) does not give a correct account of the
possible values of
, because: i) it is extended to the
low mass values sizable more than the measured points would allow it;
ii) it gives practically no chance to mass values above e.g.
MeV.
Finally there is the question of combining this result with the
other five ones of other experiments. What should we use as input
for the global analysis? Honestly, at this point
we cannot pretend to have not seen the outcome shown in
Fig. 14 and to use just the resulting average and
standard uncertainty. We also cannot feed into the model
the complicated posterior we have got. Therefore the only
solution is to make a new combined analysis, but using
all individual results of Ref. [8].
For sake of clarity all points are repeated in
Tab. 2.
Figure:
Combined analysis obtained considering the
individual points of [8]. The not trivial final
pdf of the sceptical analysis (thick continuous black line)
can be summarized as
MeV. The weighed
average (dashed red) leads instead
,
which becomes
(solid gray Gaussian
just below the dashed red one) when the standard deviation
is scaled by the factor
.
 |
The result of the analysis, plotted
in Fig 15, is quite surprising
on a first sight: while the standard weighted
average is practically the same
of Fig. 13
(small differences might be attributed to
rounding23),
the sceptical
combination moves up, disfavoring the low mass solution and
yielding
MeV.
“The same as the PDG result”, one would promptly shout at this point,
“and after so much work!”. Well, yes and no...
Indeed, the PDG numbers
were obtained considering an arbitrarily enlarged
uncertainty for the combined result of Ref. [8].
Applying, instead, the scaling prescription to
the nine individual points of Tab. 2
a value of
MeV would have been obtained,
keV below the result of the sceptical analysis
(see Fig.15).
Certainly this
bias will not harm
our understating of fundamental physics, but it is better
to avoid this kind of biases because they could
perhaps be important in other measurements.
Subsections